Weighing Balances

XPR Micro-Analytical Balances

Direct Dosing of Very Small Samples
Specification XPR36C
XPR36
XPR36DR
XPR56C
XPR56
XPR56DR
Limit values
Maximum capacity (full range/fine range)32 g32 g32 g / 8.1 g52 g52 g52 g / 11 g
Readability (full range/fine range)0.001 mg0.001 mg0.01 mg / 0.001 mg0.001 mg0.001 mg0.01 mg / 0.001 mg
Repeatability, 5% load0.001 mg0.0015 mg0.002 mg0.001 mg0.0012 mg0.002 mg
Linearity deviation0.012 mg0.012 mg0.02 mg0.012 mg0.02 mg0.03 mg
Typical values
Repeatability, 5% load0.0006 mg0.0007 mg0.001 mg0.0006 mg0.0007 mg0.001 mg
Linearity deviation0.005 mg0.005 mg0.008 mg0.005 mg0.005 mg0.008 mg
Sensitivity offset, nominal load0.06 mg0.06 mg0.08 mg0.08 mg0.08 mg0.08 mg
Minimum weight (k = 2, U = 1%)0.12 mg0.12 mg0.14 mg0.12 mg0.14 mg0.14 mg
Minimum weight (according to USP)1.2 mg1.4 mg2 mg1.2 mg1.4 mg2 mg
Setting time3.5 s3.5 s2.5 s3.5 s3.5 s2.5 s
Dimensions
Weighing pan size (mm)Ø 35; 40 × 4040 × 4040 × 40Ø 35; 40 × 4040 × 4040 × 40
Draft shield (W × D × H mm)83 × 83 × 115
Weighing pan typeSmartGrid, Hanging weighing panSmartGridSmartGridSmartGrid, Hanging weighing panSmartGridSmartGrid

* Minimum weight typical at 5% load



Features

Accurate Results Efficient Operation Quality Assurance Seamless Process
  • Active Temperature Control system (ATC™)
  • proFACT Advanced internal adjustment
  • Compact ionizer ready
  • StaticDetect™
  • 7" color touchscreen with gesture control (glove compatible)
  • ErgoClip for direct dosing
  • SmartSens for touch-less door operation
  • SmartTrac dosing guide
  • SmartGrid weighing pan
  • Built-in results notepad
  • Storage of methods
  • Automatic draft shield doors
  • GWP Approved built-in quality monitoring
  • MinWeigh warning function
  • Balance ready StatusLight
  • LevelControl: Level warning with graphical levelling guide
  • Tolerance profiles
  • User management
  • Adjustment and routine test history
  • Change history function
  • LabX ready
  • Calibry ready
  • Multiple ports for easy connectivity and data export:
    4 × USB, 1 × LAN




MX Analytical Balances

Fast and Accurate
Specification MX105 MX105DU MX205DU
Limit values
Capacity120 g120 g220 g
Readability0.01 mg0.1 mg0.1 mg
Capacity of fine range42 g82 g
Readability in fine range0.01 mg0.01 mg
Repeatability (at 5% load)0.02 mg0.02 mg0.02 mg
Linearity deviation0.1 mg0.2 mg0.2 mg
Sensitivity offset (at nominal load) 0.4 mg0.4 mg0.4 mg
Typical values
Repeatability (at 5% load)0.0125 mg0.0125 mg0.0125 mg
Linearity deviation0.06 mg0.06 mg0.06 mg
Sensitivity offset (at nominal load) 0.25 mg0.25 mg0.5 mg
Minimum weight (USP, tolerance = 0.10%)25 mg25 mg25 mg
Minimum weight (tolerance = 1%)2.5 mg2.5 mg2.5 mg
Settling time2 s2 s2 s
Dimensions and other specifications
Balance dimensions (W × D × H)222 × 379 × 353 mm
Weighing pan diameter80 mm




MR Analytical Balances

Robust and Reliable
Specification MR104 MR204 MR304
Limit values
Capacity120 g220 g320 g
Readability0.1 mg0.1 mg0.1 mg
Repeatability (at 5% load)0.1 mg0.1 mg0.1 mg
Linearity deviation0.2 mg0.2 mg0.2 mg
Sensitivity offset (at nominal load) ▲0.5 mg0.8 mg1 mg
Typical values
Repeatability (at 5% load) ▼0.08 mg0.08 mg0.08 mg
Linearity deviation0.06 mg0.06 mg0.06 mg
Sensitivity offset (at nominal load) ▲0.3 mg0.5 mg0.6 mg
Minimum weight (USP, tolerance = 0.10%) ▼160 mg160 mg160 mg
Minimum weight (tolerance = 1%) ▼16 mg16 mg16 mg
Settling time2 s2 s2 s
Dimensions and other specifications
Balance dimensions (W × D × H)209 × 351 × 354 mm
Weighing pan diameter90 mm

▲ after adjustment with internal weight
▼ determined at 5% load, k = 2
All models are available as approved versions.





MA Precision Balances Efficient and Simple

Specification MA103 MA203 MA303 MA503
Limit values
Capacity120 g220 g320 g520 g
Readability1 mg1 mg1 mg1 mg
Repeatability (at 5% load)1 mg1 mg1 mg1 mg
Linearity deviation2 mg2 mg2 mg2 mg
Sensitivity offset (at nominal load)6 mg8 mg8 mg8 mg
Typical values
Repeatability (at 5% load)0.7 mg0.7 mg0.7 mg0.7 mg
Linearity deviation0.6 mg0.6 mg0.6 mg0.6 mg
Sensitivity offset (at nominal load)4 mg5 mg5 mg5 mg
Minimum weight (USP, tolerance = 0.10%)14.4 g1.4 g1.4 g1.4 g
Minimum weight (tolerance = 1%)140 mg140 mg140 mg140 mg
Setting time1.5 s1.5 s1.5 s1.5 s
Dimensions and other specifications
Balance dimensions (W × D × H)209 × 354 × 354 mm
Weighing pan diameter120 mm


Specification MA602 MA1002 MA2002 MA3002
Limit values
Capacity620 g1.2 kg2.2 kg3.2 kg
Readability0.01 g0.01 g0.01 g0.01 g
Repeatability (at 5% load)10 mg10 mg10 mg10 mg
Linearity deviation20 mg20 mg20 mg20 mg
Sensitivity offset (at nominal load)40 mg60 mg80 mg80 mg
Typical values
Repeatability (at 5% load)7 mg7 mg7 mg7 mg
Linearity deviation6 mg6 mg6 mg6 mg
Sensitivity offset (at nominal load)25 mg40 mg50 mg50 mg
Minimum weight (USP, tolerance = 0.10%)14 g14 g14 g14 g
Minimum weight (tolerance = 1%)1.4 g1.4 g1.4 g1.4 g
Setting time1 s1 s1 s1 s
Dimensions and other specifications
Balance dimensions (W × D × H)209 × 354 × 100 mm
Weighing pan dimensions (W × D)180 × 180 mm




MA Analytical Balances Efficient and Simple

Specification MA55 MA95 MA155DU
Limit values
Capacity52 g92 g152 g
Readability0.01 mg0.01 mg0.1 mg
Capacity of fine range62 g
Readability in fine range0.01 mg
Repeatability (at 5% load)0.03 mg0.03 mg0.03 mg
Linearity deviation0.1 mg0.1 mg0.2 mg
Sensitivity offset (at nominal load)0.4 mg0.4 mg0.8 mg
Typical values
Repeatability (at 5% load)0.015 mg0.015 mg0.015 mg
Linearity deviation0.03 mg0.03 mg0.06 mg
Sensitivity offset (at nominal load)0.25 mg0.25 mg0.25 mg
Minimum weight (USP, tolerance = 0.10%)30 mg30 mg30 mg
Minimum weight (tolerance = 1%)3 mg3 mg3 mg
Setting time4 s4 s4 s
Dimensions and other specifications
Balance dimensions (W × D × H)209 × 354 × 354 mm
Weighing pan diameter80 mm


Specification MA54 MA104 MA204
Limit values
Capacity52 g120 g220 g
Readability0.1 mg0.1 mg0.1 mg
Repeatability (at 5% load)0.1 mg0.1 mg0.1 mg
Linearity deviation0.2 mg0.2 mg0.2 mg
Sensitivity offset (at nominal load)0.3 mg0.5 mg0.8 mg
Typical values
Repeatability (at 5% load)0.08 mg0.08 mg0.08 mg
Linearity deviation0.06 mg0.06 mg0.06 mg
Sensitivity offset (at nominal load)0.15 mg0.3 mg0.5 mg
Minimum weight (USP, tolerance = 0.10%)160 mg160 mg160 mg
Minimum weight (tolerance = 1%)16 mg16 mg16 mg
Setting time2 s2 s2 s
Dimensions and other specifications
Balance dimensions (W × D × H)209 × 354 × 354 mm
Weighing pan diameter90 mm




LA Balances Simple and Affordable

Specification LA84 LA104 LA204
Limit values
Capacity82 g120 g220 g
Readability0.1 mg0.1 mg0.1 mg
Repeatability (at 5% load)0.08 mg0.08 mg0.08 mg
Linearity deviation0.2 mg0.2 mg0.2 mg
Sensitivity offset (at nominal load)0.5 mg0.5 mg0.8 mg
Typical values
Repeatability (at 5% load)0.06 mg0.06 mg0.06 mg
Linearity deviation0.2 mg0.2 mg0.2 mg
Sensitivity offset (at nominal load)0.3 mg0.3 mg0.5 mg
Minimum weight (USP, tolerance = 0.10%)160 mg160 mg160 mg
Minimum weight (tolerance = 1%)16 mg16 mg16 mg
Setting time3 s3 s3 s
Dimensions and other specifications
Balance dimensions (W × D × H)210 × 308 × 345 mm
Weighing pan diameter80 mm


Specification LA203 LA403 LA2002 LA4002
Limit values
Capacity220 g420 g2200 g4200 g
Readability1 mg1 mg10 mg10 mg
Repeatability (at 5% load)1 mg1 mg10 mg10 mg
Linearity deviation2 mg2 mg20 mg20 mg
Sensitivity offset (at nominal load)3 mg3 mg8 mg8 mg
Typical values
Repeatability (at 5% load)0.7 mg0.7 mg7 mg7 mg
Linearity deviation1 mg1 mg6 mg6 mg
Sensitivity offset (at nominal load)1.4 mg1.4 mg4 mg4 mg
Minimum weight (USP, tolerance = 0.10%)14 mg14 mg144 mg144 mg
Minimum weight (tolerance = 1%)1.4 g1.4 g14.4 g14.4 g
Dimensions and other specifications
Balance dimensions (W × D × H)210 × 308 × 280 mm200 × 308 × 102 mm
Weighing pan diameter100 mm180 mm




Footer
CATALOGUE