Email : sales@aadarshtech.com
+91 - 9619519171 / +91 - 8691007010

Products

Our products have been successfully used various fields. The company’s global footprint and the critical nature of its products and services play a key role in expanding its distribution to many countries across the world.

Testing Solution For Facial Mask

GB-KF30010 Face Mask Particulate Filtration Efficiency Tester

GB-KF30010 Face Mask Particulate Filtration Efficiency Tester

GBPI offers mask testing machines for surgical face masks, N95 masks, Melt-blown fabric. BFE Tester, PFE Tester, N95 Tester, Melt-blown fabric Tester

Application:

It is to test the filtering efficiency of particulate protective masks and medical masks and to determine the resistance of ordinary fabrics and medical protective masks to the constant flow of airflow.

Standards

GB/T 32610-2016、GB 2626-2006、GB 19082-2009、GB 19083-2010
GB 24539-2009 、YY 0469-2011 、YYT 0969-2013

Features

  • Using a cold aerosol generator to produce continuous and stable aerosol particles, convenient for filling the solution.
  • Measurement of aerosol concentration using a high-precision PM2.5 sensor. Anti-leakage design of whole particles to protect 2 sets of Aerosol generator:Salty aerosol generator and oily aerosol generator. the safety of laboratory personnel.
  • Equipped with aerosol particle neutralization device.
  • The pneumatic clamp is equipped with a protection device, which is safe and convenient to use.
  • Configure temperature and humidity sensor, real-time display of ambient temperature and humidity(Temperature and humidity requirements:25℃±5℃,30%RH±10%RH).
    Equipped with glass rotor flowmeter, vacuum pump, laser dust particle counter.
  • Control system: The computer controls the test process, automatically collects data, and configures a special computer and test software.
  • The computer automatically tests the gas concentration and calculates the filtration efficiency. It can save, output, query and print the test data.
Item specification
Test range 0-99.999%
Flowmeter range (10-100) L/min
Sampling frequency 1-9999/min
Particle concentration (0.001-200) mg/m3
Counting median diameter Salt particles(0.075±0.02)、(0.185±0.02)μm
Geometric standard deviation of particle size distribution Salt particles≤1.86、 Oil particles≤1.60
Dynamic detection range 0.001-100 mg/m3, accuracy1%
Dimension 460mm x 525mm x 1430mm
Power supply AC 220V, 50Hz
Differential pressure sensor range 0~500Pa